Conference Abstract | Volume 8, Abstract ELIC2025238 (Oral 034) | Published: 19 Aug 2025
Nsonghomanyi Fritz Roland Fonkeng1,&, Hanesh Fru Chi1, Adama Ahmad2, Shamzu Munzali2, Ifeanyi Nwafor3, Emmanuel Agogo1
1FIND, Geneva, Switzerland, 2Nigeria Centre for Disease Control and Prevention, Abuja, Nigeria, 3Alex Ekwueme Federal University Teaching Hospital Lassa Fever Laboratory, Abakiliki, Nigeria
&Corresponding author: Nsonghomanyi Fritz Roland Fonkeng, FIND Geneva, Switzerland, Email: fritz.fonkeng@finddx.org
Received: 11 May 2025, Accepted: 09 Jul 2025, Published: 19 Aug 2025
Domain: Infectious Disease Epidemiology
Keywords: External quality assurance, Lassa fever, diagnostic accuracy, laboratory quality, Nigeria, CT values
©Nsonghomanyi Fritz Roland Fonkeng et al. Journal of Interventional Epidemiology and Public Health (ISSN: 2664-2824). This is an Open Access article distributed under the terms of the Creative Commons Attribution International 4.0 License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Cite this article: Nsonghomanyi Fritz Roland Fonkeng et al. Establishing an external quality assurance (EQA) system for Lassa fever testing in Nigeria. Journal of Interventional Epidemiology and Public Health. 2025;8(ConfProc5):00034. https://doi.org/10.37432/jieph-confpro5-00034
Lassa fever is a viral hemorrhagic disease endemic in West Africa, capable of progressing from mild illness to multiorgan failure. During the 2018 outbreak in Nigeria, 423 confirmed cases were recorded with a case fatality rate of 25%. Given the absence of commercial diagnostic assays for Lassa virus, ensuring test accuracy through quality assurance systems is critical. External Quality Assurance (EQA) programs serve to evaluate and enhance the reliability of laboratory diagnostics by providing standardized, blinded testing materials.
An EQA program was initiated using Lassa fever-positive samples with known cycle threshold (CT) values. These samples were pooled, heat-inactivated, blinded, and distributed to all active testing laboratories. Laboratories conducted testing according to their routine procedures. Concordance between expected and reported results was used to evaluate test performance.
In the first round, 16 panels were distributed and 13 tested, yielding a 92% concordance rate. In the second round, all 15 distributed panels were tested, achieving a 100% concordance rate. These results indicate both progress in testing quality and the responsiveness of labs to continuous performance monitoring.
The EQA system has proven effective in benchmarking diagnostic quality and identifying areas for capacity strengthening. Annual implementation of this program will support ongoing quality improvement, guide targeted training interventions, and serve as a tool to evaluate the readiness of new laboratories for Lassa fever testing.